The Majorana Demonstrator comprises two arrays of high-purity germanium detectors constructed to search for neutrinoless double-beta decay in 76Ge and other physics beyond the Standard Model. Its readout electronics were designed to have low electronic noise, and radioactive backgrounds were minimized by using low-mass components and low-radioactivity materials near the detectors. This paper provides a description of all components of the Majorana Demonstrator readout electronics, spanning the front-end electronics and internal cabling, back-end electronics, digitizer, and power supplies, along with the grounding scheme. The spectroscopic performance achieved with these readout electronics is also demonstrated.

Digital Object Identifier (DOI)

https://doi.org/10.1088/1748-0221/17/05/t05003

Rights

©2022 The Author(s). Published by IOP Publishing Ltd on behalf of Sissa Medialab. Original content from this work may be used under the terms of the Creative Commons Attribution 4.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

APA Citation

N. Abgrall, Amman, M., Arnquist, I. J., Avignone, F. T., Barabash, A. S., Barton, C. J., Barton, P. J., Bertrand, F. E., Bhimani, K. H., Bos, B., Bradley, A. W., Burritt, T. H., Busch, M., Buuck, M., Caldwell, T. S., Chan, Y-D., Christofferson, C. D., Chu, P.-H., Clark, M. L., & Cooper, R. J. (2022). The Majorana Demonstrator readout electronics system. Journal of Instrumentation, 17(05), T05003–T05003. https://doi.org/10.1088/1748-0221/17/05/t05003